VNG Infant ICU Incubator 3000
Contact our team for pricing, configuration options and availability.
Thanks! Our team will contact you shortly.
Closed-care NICU incubator with air- and skin-mode servo temperature control, plus optional humidity, oxygen and weighing modules.
An ambient-temperature probe compares ambient and set temperature, adjusting heating output for a stable inside-hood microclimate. By VNG Medical.
At a glance
- Air- and skin-mode servo temperature control for stable microclimate
- Optional servo humidity control (30–95% RH) with 24-hour reservoir capacity
- Optional servo oxygen control (21–65% FiO2)
- Optional integrated weighing scale (300g–8kg, 1g resolution)
- Low operating noise (≤45dBA) and low CO₂ build-up (<0.5%)
- Tiltable soft-bed mattress (±12°) for positioning
OVERVIEW
The VNG Infant Incubator 3000 is a closed-care NICU incubator. An ambient temperature probe compares the gap between ambient and set temperature and adjusts heating power output accordingly, to provide a more stable inside-hood air temperature for the infant. Optional servo-controlled humidity and oxygen modules, and a built-in weighing scale option, let it be configured for higher-acuity NICU care.
TEMPERATURE CONTROL
- Air mode control range: 20°C–37°C, accuracy ≤±1.0°C
- Skin mode control range: 34°C–37°C, accuracy ±0.3°C
- Temperature rise time at 22°C ambient: under 20 minutes
- Temperature uniformity: 0.8°C; variability: <0.2°C
OPTIONAL HUMIDITY CONTROL (SERVO)
- Control range: 30%–95% RH, accuracy ±5% RH
- Reservoir capacity: 1500ml, up to 24 hours' operation without refilling
OPTIONAL OXYGEN CONTROL (SERVO)
- Control range: 21% to 65%
- Display accuracy: ±2%; control accuracy at 100% calibration: ±3%
OPTIONAL WEIGHING SCALE
- Weight range: 300g to 8kg
- Display resolution: 1g; accuracy: ±10g
OTHER SPECIFICATIONS
- Soft bed mattress: 73.6 × 38.6 × 1.8cm, tilt ±12°
- Air flow velocity inside hood: <10cm/sec
- Operating noise level: ≤45dBA
- CO² level: <0.5%
- Operating environment: 20°C–30°C, 10–95% RH
Frequently asked questions
What kind of incubator is the VNG Infant Incubator 3000?
How does it maintain a stable temperature?
Can it control humidity and oxygen?
Does it include a weighing scale?
How noisy is it inside the hood?
What's the mattress tilt range?
Related reading
Understanding the Role of Anesthesia in Pediatric Surgery: Best Practices and Equipment
This article delves into pediatric anesthesia's critical role in surgery, emphasizing best practices and the necessary equipment to ensure safety in NICU settings.
Read article Liver/HepatologyThe Role of Diagnostic Equipment in Early Detection of Pediatric Liver Conditions
This article explores the critical role of diagnostic equipment in the early detection of pediatric liver conditions, emphasizing its impact on patient care.
Read article