OVERVIEW
The VNG Infant Incubator 3000 is a closed-care NICU incubator. An ambient temperature probe compares the gap between ambient and set temperature and adjusts heating power output accordingly, to provide a more stable inside-hood air temperature for the infant. Optional servo-controlled humidity and oxygen modules, and a built-in weighing scale option, let it be configured for higher-acuity NICU care.
TEMPERATURE CONTROL
- Air mode control range: 20°C–37°C, accuracy ≤±1.0°C
- Skin mode control range: 34°C–37°C, accuracy ±0.3°C
- Temperature rise time at 22°C ambient: under 20 minutes
- Temperature uniformity: 0.8°C; variability: <0.2°C
OPTIONAL HUMIDITY CONTROL (SERVO)
- Control range: 30%–95% RH, accuracy ±5% RH
- Reservoir capacity: 1500ml, up to 24 hours' operation without refilling
OPTIONAL OXYGEN CONTROL (SERVO)
- Control range: 21% to 65%
- Display accuracy: ±2%; control accuracy at 100% calibration: ±3%
OPTIONAL WEIGHING SCALE
- Weight range: 300g to 8kg
- Display resolution: 1g; accuracy: ±10g
OTHER SPECIFICATIONS
- Soft bed mattress: 73.6 × 38.6 × 1.8cm, tilt ±12°
- Air flow velocity inside hood: <10cm/sec
- Operating noise level: ≤45dBA
- CO² level: <0.5%
- Operating environment: 20°C–30°C, 10–95% RH